DS-800 Calibration instruction

 

1.  Switch on the scale.  Wait for 0.0 to be displayed.

2.  Press and hold the UNIT button for 3 seconds, to enter the calibration mode.

The LCD displays    CAL  ,  starting the automatic calibration.

2. LCD displays   LoAd   à  200 (lb).

3. Load 200 lb on the platform then press UNIT button to confirm.

4. LCD displays 200.0 lb.  The calibration is complete.

5. Remove the 200 lb from the platform to make sure it returns to "0.0"

6. Place the 200 lb back on the platform to make sure it reads "200.0 lb"

 

 

 