Calibration instruction 1.
Switch on the scale. Wait for 0.0 to be displayed. 2.
Press and hold the “UNIT” button for 3 seconds, to enter the calibration mode. The LCD displays 2. LCD displays LoAd à 200
(lb). 3. Load 200 lb on the platform then press “UNIT” button to confirm. 4. LCD displays “200.0 lb”. The calibration is complete. 5. Remove the 200 lb from the platform to make sure it returns to "0.0" 6.
Place the 200 lb back on the platform to make sure it reads "200.0 lb"
|